| Name: | Door opening tests on upright vaccine refrigerator & assessment of thermal shock |
| Description: | Tests were performed using the True Energy Sure Chill® Vaccine Refrigerator in order to assess the extent of any thermal shock to vaccines on opening the refrigerator door at high ambient temperatures. Following a protocol which mimicked usage in the field, results showed that there was no thermal shock to vaccines stored in the Sure Chill® vaccine refrigerator when the refrigerator door was opened to 90o for 3 minutes at an ambient temperature of 43oC. |